Gravity on the move

 

 

 

 

 

 

 

 

 

Home
Up

Lab Test 3

Scorsby Bench

Test Description

Date                                June 2001

Location                           Gravimetric Technologies labs, Moscow

Test description                Three series of tests carried out,  each test

                                       lasting approximately 1 hr 40 minutes

Oscillation periods            25,15, and 7 seconds typical

Duration of tests               8 hours total

 

Results

Filter                                Various filters applied as indicated in the tables below

                                                              

Notes:                             Gravimeter noise in bandwidth 0 to 0.017 Hz after warming:

·          without motion 0.07 mGal

·          with motion     0.17 mGal

                                                   More detailed results are given in the tables below.

                                                   Inertial errors from Scorsby motion not compensated.

                                                   As the gravimeter warmed up over the 8 hours of the

                                                   tests, the RMS noise decreased from 0.18 to 0.07

                                                   mGals with a 60 second filter (static) and from 0.19

                                                   to 0.17 mGals (moving).

 

         RMS while standing (mGal)

Warming up

Filtering time

2 sec

20 sec

60 sec

100 sec

200 sec

G29 0833 DAT

0.2496

0.2050

0.1812

0.1702

0.1547

G291201  DAT

0.2049

0.1415

0.1023

0.0865

0.0632

G291219  DAT

0.1934

0.1298

0.0894

0.0770

0.0674

G291413  DAT

0.1934

0.1298

0.0894

0.0770

0.0674

G291609  DAT

0.1576

0.1039

0.0739

0.0617

0.0451

 

         RMS while rolling (mGal)

Filtering time

2 sec

20 sec

60 sec

100 sec

200 sec

G291015 DAT

54.6652

5.4051

0.1869

0.1649

0.1485

G291238 DAT

125.2067

0.9875

0.1811

0.1676

0.1553

G291431 DAT 

268.9006

0.1995

0.1699

0.1600

0.1463

 

Warming up

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Download case history in pdf format (120 KB).

 